Purpose:  The purpose of the Honan Intraocular Pressure Reducer is to provide increased safety for intraocular
surgeries.

What does it do?
Controlled preoperative ocular compression creates a lower pressure inside the eye (a soft eye) by reducing the
fluid volume of the vitreous and other tissues inside and behind the eye.

How does a softer eye increase safety in surgery?
  • A softer eye reduces the tendency of the contents of the eye such as vitreous humor or iris to bulge forward
    when the eye is opened for surgery.
  • The eye has a “normal” pressure inside to maintain its shape.  At the moment of a surgical incision the  
    pressure inside the eye plunges suddenly down to that of atmospheric pressure.  A layer of choroidal vessels
    lies just outside the inside retinal layer of the wall of the eyeball.  Healthy resilient choroidal vessels are able
    to withstand the shock of the sudden drop in eye pressure.  Rarely fragile aging choroidal vessels can crack
    and bleed when the eye pressure suddenly drops.  The sudden bleeding can cause a devastating internal
    expulsive hemorrhage that can force the eye contents out of the surgical incision.  Preoperative ocular
    compression can lower the intraocular pressure, which will reduce the shock on the choroidal vessels at the
    time of surgical incision and reduce the risk of hemorrhage.

Method of use
  • Tape the eyelids closed to avoid corneal abrasion.
  • Apply a cotton eye pad or 4 x 4 gauze over the eye
  • Apply the head strap around the back of the head
  • Position the balloon over the cotton pad on the eye to be operated and secure the headband firmly but not
    tight
  • Inflate the balloon to the desired pressure.  Some use 30-mm mercury.  A pressure of 20-mm mercury is
    satisfactory.
  • Keep the balloon on the eye 20 minutes or longer
  • The balloon may be removed to apply eye drops and reapplied
  • Remove the balloon in the OR just before the surgical prep.
  • Early removal of the balloon will allow the normal intraocular pressure to return and increase the risk of
    hemorrhage.